Visit
Official Website

Testo 0635 1052 | Fume cupboard probe (digital) - wired

Previous 7 / 17 Next

Quantity
Add Quotation


Testo 0635 1052 | Fume cupboard probe (digital) - wired

Product Description

Use the fume cupboard probe (hot wire probe) with the compatible testo multifunction measuring instrument (please order separately) to determine air velocity, volume flow and air temperature. It is particularly suitable for low air velocities up to +5 m/s.

 

Fume cupboard probe – features and areas of application

Fume cupboard testing plays an important role in ensuring the safety of those who work with substances that are harmful to health. For example, the fume cupboard offers protection against explosions as well as hazardous gases and vapours.

Use the fume cupboard probe to measure the inflow rate and exhaust air volume flow in laboratory fume cupboards in accordance with DIN EN 14175-3/-4. This will guarantee that the fume cupboard is set up correctly and can function properly for your customer.

Use the fixed cable to connect the fume cupboard probe to the measuring instrument (please order separately). The clearly structured measurement menu for volume flow makes the measuring instrument intuitive to operate. The volume flow is precisely calculated thanks to convenient input of the size and geometry of the duct cross-section. Timed and multi-point mean calculation, average volume flow, current reading and min./max. values are displayed in the measuring instrument. To ensure accurate measurement results, air density compensation is carried out via the integrated absolute pressure measurement.

 

Intelligent calibration concept

You will get exceptionally accurate measurement results with the digital probe, because the measuring instrument makes measurement uncertainty a thing of the past. You only need to send the probe in for calibration – so the measuring instrument remains in continuous use.
 


Delivery Scope

Fume cupboard probe with fixed cable (length 1.4 m) including calibration protocol.

Techinical Data

Temperature - NTC
Measuring range 0 to +50 °C
Accuracy ±0.5 °C
Resolution 0.1 °C
 
Absolute Pressure
Measuring range +700 to +1100 hPa
Accuracy ±3.0 hPa
Resolution 0.1 hPa
 
Velocity - Hot wire
Measuring range 0 to 5 m/s
Accuracy ±(0.02 m/s + 5 % of mv)
Resolution 0.01 m/s
 
General technical data
Storage temperature -20 to +60 °C
Dimensions 350 x 50 x 40 mm
Operating temperature 0 to +50 °C
Cable length 1.4 m
Standards EN 14175 3/ 4
Weight 230 g


 
Please leave your enquiry here, we will reply as soon as possible.
Name*  
Company Name  
Product Interested  
Quantity  
Email*  
Contact No.*  
Messages*  
   

You have 0 items in you cart. Would you like to checkout now?
0 items
Recent Updates
Subscribe Newsletter