Please do not use the probe head in condensing atmospheres. For continuous application in high-humidity ranges
> 80% RH at ≤ 30 °C for > 12 h
> 60% RH at > 30 °C for > 12 h
please contact Testo Service or contact us via the Testo website
Technical Data
Temperature - NTC | |
---|---|
Measuring range | 0 to +50 °C |
Accuracy | ±0.5 °C |
Resolution | 0.1 °C |
Humidity - Capacitive | |
---|---|
Measuring range | 5 to 95 %RH |
Accuracy |
±3 %RH (10 to 35 %RH) ±2 %RH (35 to 65 %RH) ±3 %RH (65 to 90 %RH) ±5 %RH (Remaining Range) ±0.06 %RH/K (0 to +50 °C) |
Resolution | 0.1 %RH |
Absolute Pressure | |
---|---|
Measuring range | +700 to +1100 hPa |
Accuracy | ±3.0 hPa |
Resolution | 0.1 hPa |
Ambient CO₂ | |
---|---|
Measuring range | 0 to 10000 ppm |
Accuracy |
±(50 ppm + 3 % of mv) (0 to 5000 ppm) ±(100 ppm + 5 % of mv) (5001 to 10000 ppm) |
Resolution | 1 ppm |
General technical data | |
---|---|
Storage temperature | 0 to +50 °C |
Weight | 60 g |
Dimensions | 165 x 30 x 30 mm |
Operating temperature | 0 to +50 °C |
Length probe shaft | 130 mm |
Probe head diameter | 30 mm |
Product colour | Black |